Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Metrix MX 51: (1 year) CAL VER /5520 INSTRUMENT: Metrix MX 52: (1 year) CAL VER /5520 DATE: 04-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 72 NUMBER OF LINES: 278 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke Corp. makes no warranty, expressed or implied, # as to the fitness or useability of this procedure. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD INITIAL SETUP 1.007 DISP Turn on the UUT. 1.008 OPBR Is the battery indicator illuminated? 1.009 JMPF 1.011 1.010 DISP Replace battery before proceeding. 1.011 HEAD {AC VOLTAGE} 1.012 DISP Rotate the Function Switch to V AC. 1.012 DISP Press "RANGE" to lock the 5V Range. 1.013 DISP Connect the 5520A to the UUT as follows: 1.013 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 1.013 DISP [32] 5520A NORMAL LO to UUT COM 1.014 JMPT 9.002 Metrix MX 52: (1 year) CAL VER /5520 # MX 51 1.015 HEAD AC VOLTAGE: {5V Range} 1.016 5520 5 4.900V 0.75% 0.002U 60H SI 2W 2.001 5520 5 4.900V 1.7% 0.002U 1kH SI 2W 3.001 HEAD AC VOLTAGE: {50V Range} 3.002 DISP Press "RANGE" to lock the 50V Range. 3.003 5520 50 49.00V 0.75% 0.02U 60H SI 2W 4.001 5520 50 49.00V 1.7% 0.02U 1kH SI 2W 5.001 HEAD AC VOLTAGE: {500V Range} 5.002 DISP Press "RANGE" to lock the 500V Range. 5.003 5520 500 490.0V 0.75% 0.2U 60H SI 2W 6.001 5520 500 490.0V 1.7% 0.2U 1kH SI 2W 7.001 HEAD AC VOLTAGE: {750V Range} 7.002 DISP Press "RANGE" to lock the 750V Range. 7.003 5520 750 750V 0.75% 2U 60H SI 2W 8.001 5520 750 750V 1.7% 2U 1kH SI 2W 9.001 JMP 19.001 # MX 52 9.002 HEAD AC VOLTAGE: {5V Range} 9.003 5520 5 4.900V 0.75% 0.002U 60H SI 2W 10.001 5520 5 4.900V 3% 0.002U 20kH SI 2W 11.001 HEAD AC VOLTAGE: {50V Range} 11.002 DISP Press "RANGE" to lock the 50V Range. 11.003 5520 50 49.00V 0.75% 0.02U 60H SI 2W 12.001 5520 50 49.00V 3% 0.02U 20kH SI 2W 13.001 HEAD AC VOLTAGE: {500V Range} 13.002 DISP Press "RANGE" to lock the 500V Range. 13.003 5520 500 490.0V 0.75% 0.2U 60H SI 2W 14.001 5520 500 490.0V 1.7% 0.2U 1kH SI 2W 15.001 5520 500 490.0V 3% 0.2U 5kH SI 2W 16.001 5520 500 490.0V 3% 0.2U 10kH SI 2W 17.001 HEAD AC VOLTAGE: {750V Range} 17.002 DISP Press "RANGE" to lock the 750V Range. 17.003 5520 750 750V 0.75% 2U 60H SI 2W 18.001 5520 750 750V 1.75% 2U 1kH SI 2W 19.001 HEAD AC VOLTAGE: {500mV Range} 19.002 DISP Rotate the Function Switch to mV. 19.003 JMPT 21.002 Metrix MX 52: (1 year) CAL VER /5520 # MX 51 19.004 5520 500 490.0mV 0.75% 0.2U 60H SI 2W 20.001 5520 500 490.0mV 1.7% 0.2U 1kH SI 2W 21.001 JMP 23.001 # MX 52 21.002 5520 500 490.0mV 0.75% 0.2U 60H SI 2W 22.001 5520 500 490.0mV 2% 0.2U 1kH SI 2W 23.001 HEAD {DC VOLTAGE} 23.002 HEAD DC VOLTAGE: {500mV Range} 23.003 DISP Press "AC/DC" to select mV DC. 23.004 5520 500 490.0mV 0.1% 0.1U 2W 24.001 HEAD DC VOLTAGE: {5V Range} 24.002 DISP Rotate the Function Switch to V DC. 24.002 DISP Press "RANGE" to lock the 5V Range. 24.003 5520 5 4.900V 0.1% 0.001U 2W 25.001 HEAD DC VOLTAGE: {50V Range} 25.002 DISP Press "RANGE" to lock the 50V Range. 25.003 5520 50 49.00V 0.1% 0.01U 2W 26.001 HEAD DC VOLTAGE: {500V Range} 26.002 DISP Press "RANGE" to lock the 500V Range. 26.003 5520 500 490.0V 0.1% 0.1U 2W 27.001 HEAD DC VOLTAGE: {1000V Range} 27.002 DISP Press "RANGE" to lock the 1000V Range. 27.003 5520 1000 1000V 0.1% 1U 2W 28.001 HEAD {LOGIC TESTS} 28.002 DISP Rotate the Function Switch to LOGIC. 28.003 EVAL Does the UUT indicate an {OPEN}? 29.001 5520 1.2V S 2W 29.002 EVAL Does the UUT indicate {LOGIC 0 (LO)}? 30.001 5520 1.4V S 2W 30.002 EVAL Does the UUT indicate {LOGIC 1 (HI)}? 31.001 5520 21V S 2W 31.002 EVAL Does the UUT indicate {OVERLOAD (OL)}? 32.001 HEAD {DIODE CHECK} 32.002 DISP Rotate the Function Switch to DIODE. 32.003 5520 0.5V 0.1U 2W 33.001 HEAD {RESISTANCE} 33.002 HEAD RESISTANCE: {500 kOhm Range} 33.003 DISP Rotate the Function Switch to Ohms. 33.003 DISP Press "RANGE" five times to lock the 500 kOhm Range. 33.004 5520 500 490.0kZ 0.3% 0.2U 2W 34.001 HEAD RESISTANCE: {5 MOhm Range} 34.002 DISP Press "RANGE" to lock the 5 MOhm Range. 34.003 5520 5 4.900MZ 0.5% 0.002U 2W 35.001 HEAD RESISTANCE: {40 MOhm Range} 35.002 DISP Press "RANGE" to lock the 40 MOhm Range. 35.003 5520 40 39.00MZ 2% 0.02U 2W 36.001 HEAD RESISTANCE: {500 Ohm Range} 36.002 DISP Press "RANGE" to lock the 500 Ohm Range. 36.003 DISP Connect the 5520A to the UUT as follows: 36.003 DISP [32] 5520A AUX HI to UUT V/Ohm/Diode 36.003 DISP [32] 5520A AUX LO to UUT COM 36.003 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 36.003 DISP [32] 5520A NORMAL LO to UUT COM 36.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 36.003 DISP the NORMAL leads to the SENSE leads at the UUT. 36.004 5520 500 490.0Z 0.3% 0.2U CW 37.001 HEAD RESISTANCE: {5 kOhm Range} 37.002 DISP Press "RANGE" to lock the 5 kOhm Range. 37.003 5520 5 4.900kZ 0.3% 0.002U CW 38.001 HEAD RESISTANCE: {50 kOhm Range} 38.002 DISP Press "RANGE" to lock the 50 kOhm Range. 38.003 5520 50 49.00kZ 0.3% 0.02U CW 39.001 HEAD {CONTINUITY TEST} 39.002 DISP Press "AC/DC" to select Continuity. 39.002 DISP A musical note should appear on the display. 39.003 5520 10Z S CW 39.004 EVAL Is the continuity buzzer activated? 40.001 JMPT 46.001 Metrix MX 51: (1 year) CAL VER /5520 # MX 52 only 40.002 HEAD {FREQUENCY} 40.003 DISP Rotate the Function Switch to Frequency. 40.004 HEAD FREQUENCY: {10Hz Range} 40.005 5520 10 10.0H 0.05% 0.1U 1V SI 2W 41.001 HEAD FREQUENCY: {500Hz Range} 41.002 5520 500 500.0H 0.05% 0.1U 1V SI 2W 42.001 HEAD FREQUENCY: {5kHz Range} 42.002 5520 5 5.000kH 0.05% 0.001U 1V SI 2W 43.001 HEAD FREQUENCY: {50kHz Range} 43.002 5520 50 50.00kH 0.05% 0.01U 1V SI 2W 44.001 HEAD FREQUENCY: {500kHz Range} 44.002 5520 500 500.0kH 0.05% 0.1U 1V SI 2W 45.001 HEAD {dB} 45.002 DISP Rotate the Function Switch to dB. 45.003 5520 0.0D 0.3U 1kH SI 2W 46.001 HEAD {DC HIGH CURRENT} 46.002 DISP Rotate the Function Switch to A. 46.002 DISP Press "RANGE" to lock the 5A Range. 46.003 DISP Connect the 5520A to the UUT as follows: 46.003 DISP [32] 5520A 20A to UUT 10A 46.003 DISP [32] 5520A AUX LO to UUT COM 46.004 HEAD DC HIGH CURRENT: {5A Range} 46.005 5520 5 4.900A 1.2% 0.002U HC 2W 47.001 HEAD DC HIGH CURRENT: {10A Range} 47.002 DISP Press "RANGE" to lock the 10A Range. 47.003 5520 10 9.90A 1% 0.01U HC 2W 48.001 HEAD {AC HIGH CURRENT} 48.002 DISP Press "AC/DC" to select A AC. 48.002 DISP Press "RANGE" to lock the 5A Range. 48.003 JMPT 52.002 Metrix MX 52: (1 year) CAL VER /5520 # MX 51 48.004 HEAD AC HIGH CURRENT: {5A Range} 48.005 5520 5 4.900A 1.5% 0.002U 60H SI HC 2W 49.001 5520 5 4.900A 2.5% 0.002U 1kH SI HC 2W 50.001 HEAD AC HIGH CURRENT: {10A Range} 50.002 DISP Press "RANGE" to lock the 10A Range. 50.003 5520 10 9.90A 1.5% 0.02U 60H SI HC 2W 51.001 5520 10 9.90A 2.5% 0.02U 1kH SI HC 2W 52.001 JMP 56.001 # MX 52 52.002 HEAD AC HIGH CURRENT: {5A Range} 52.003 5520 5 4.900A 1.5% 0.002U 60H SI HC 2W 53.001 5520 5 4.900A 3% 0.002U 1kH SI HC 2W 54.001 HEAD AC HIGH CURRENT: {10A Range} 54.002 DISP Press "RANGE" to lock the 10A Range. 54.003 5520 10 9.90A 1.5% 0.02U 60H SI HC 2W 55.001 5520 10 9.90A 3% 0.02U 1kH SI HC 2W 56.001 HEAD {AC LOW CURRENT} 56.002 DISP Rotate the Function Switch to mA. 56.003 DISP Connect the 5520A to the UUT as follows: 56.003 DISP [32] 5520A AUX HI to UUT uA/mA 56.003 DISP [32] 5520A AUX LO to UUT COM 56.004 HEAD AC LOW CURRENT: {500mA Range} 56.005 5520 500 490.0mA 1.5% 0.2U 60H SI 2W 57.001 5520 500 490.0mA 2.5% 0.2U 1kH SI 2W 58.001 JMPF 59.001 Metrix MX 52: (1 year) CAL VER /5520 # MX 52 only 58.002 5520 500 490.0mA 3% 0.2U 5kH SI 2W 59.001 HEAD AC LOW CURRENT: {50mA Range} 59.002 DISP Press "RANGE" to select the 50mA Range. 59.003 5520 50 49.00mA 1.5% 0.02U 60H SI 2W 60.001 5520 50 49.00mA 2.5% 0.02U 1kH SI 2W 61.001 JMPF 62.001 Metrix MX 52: (1 year) CAL VER /5520 # MX 52 only 61.002 5520 50 49.00mA 3% 0.02U 5kH SI 2W 62.001 HEAD {DC LOW CURRENT} 62.002 HEAD DC LOW CURRENT: {50mA Range} 62.003 DISP Press "AC/DC" to select mA DC. 62.003 DISP Press "RANGE" to lock the 50mA Range. 62.004 5520 50 49.00mA 1.2% 0.02U 2W 63.001 HEAD DC LOW CURRENT: {500mA Range} 63.002 DISP Press "RANGE" to lock the 500mA Range. 63.003 5520 500 490.0mA 1% 0.1U 2W 64.001 HEAD DC LOW CURRENT: {5mA Range} 64.002 DISP Rotate the Function Switch to 5mA/500uA. 64.003 5520 50 4.900mA 1% 0.001U 2W 65.001 HEAD DC LOW CURRENT: {500uA Range} 65.002 DISP Press "RANGE" to lock the 500uA Range. 65.003 5520 490.0uA 1.2% 0.2U 2W 66.001 HEAD {AC LOW CURRENT} 66.002 HEAD AC LOW CURRENT: {500uA Range} 66.003 DISP Press "AC/DC" to select uA AC. 66.003 DISP Press "RANGE" to lock the 500uA Range. 66.004 5520 500 490.0uA 1.5% 0.2U 60H SI 2W 67.001 5520 500 490.0uA 2.5% 0.2U 1kH SI 2W 68.001 JMPF 69.001 Metrix MX 52: (1 year) CAL VER /5520 # MX 52 only 68.002 5520 500 490.0uA 3% 0.2U 5kH SI 2W 69.001 HEAD AC LOW CURRENT: {5mA Range} 69.002 DISP Press "RANGE" to lock the 5mA Range. 69.003 5520 50 4.900mA 1.5% 0.002U 60H SI 2W 70.001 5520 50 4.900mA 2.5% 0.002U 1kH SI 2W 71.001 JMPF 72.001 Metrix MX 52: (1 year) CAL VER /5520 # MX 52 only 71.002 5520 50 4.900mA 3% 0.002U 5kH SI 2W 72.001 END